Data Collection fails immediately when connecting a PASPort sensor to the computer via the GLX.

Affected Products:

PS-2002 Xplorer GLX

Problem/Symptom:
Data Collection fails immediately when connecting a PASPort sensor to the computer via the GLX.

PASCO Solution:

This effects XPlorer GLXs with firmware revision 1.2.4, which is installed by DataStudio 1.9.8.2, when PASPort sensors are moved from one port to another.

  1. Close DataStudio.
  2. Copy GLX.SREC to the C:\Program Files\DataStudio\Plugins\DataSheets directory on a PC or to the /Applications/DataStudio.sys/Contents/Plugins/DataSheets/ directory on a Mac, replacing existing file. (On a Mac press the CONTROL key while clicking on the DataStudio folder and select Open Package Contents.)
  3. Restart DataStudio.
  4. When prompted to update the firmware, select "YES".

Creation Date: 06/4/2007
Last Modified: 10/27/2009