Tech Note 511 Detail

PS-2107 reads about 2x higher than it should.

Affected Products:

PS-2107 PASPORT Absolute Pressure Sensor

Problem/Symptom:
Absolute Pressure sensor reads ~2x higher than it should.

PASCO Solution:

Most units manufactured between July and October of 2004 were calibrated incorrectly. At typical sea level atmospheric pressure the units should read about 100 kPa; however, improperly functioning units will read about 200 kPa at sea level.

Please contact Technical Support to return the units for factory re-calibration.

Creation Date: 10/26/2004
Last Modified: 12/1/2004