Tech Note 992 Detail

Screen requires too much pressure to activate icons.

Affected Products:

PS-2008 SPARK Science Learning System

Problem/Symptom:
Activating icons requires too much pressure or the icons will not activate at all.

PASCO Solution:

Navigate to the home screen on the SPARK. Press and hold the two buttons to the upper right and left-hand sides of the screen. While holding these buttons, tap the screen three times. After releasing the side buttons, this will launch a screen calibration dialog. Follow the instructions by pressing the insides of the green squares.

If you are running a firmware below 1.10, please update the firmware on your SPARK SLS.

Creation Date: 02/12/2009
Last Modified: 04/26/2010