Tech Note 617 Detail

Sampling errors occur when using the PS-2159 Digital Adapter with the CI-6538 or CI-6742 and the PS-2100A USB Link.

Affected Products:

PS-2159 PASPORT Digital Adapter
CI-6742 Motion Sensor II
CI-6538 Rotary Motion Sensor

Problem/Symptom:
Sampling errors occur when using the PS-2159 Digital Adapter with the CI-6538 Rotary Motion Sensor or CI-6742 Motion Sensor and the PS-2100A USB Link.

PASCO Solution:
  1. Close DataStudio.
  2. If you do not already have DataStudio 1.9.8.2 installed, please download and install it from here.
  3. Delete c:/Program Files/DataStudio/Plugins/DataSheets/UsbLink2.hex.
  4. Copy USBLink2.hex to C:\Program Files\DataStudio\Plugins\DataSheets.
  5. Restart DataStudio.
  6. Allow DataStudio to update the firmware, when prompted.

Creation Date: 10/17/2007
Last Modified: 10/27/2009