Calibration does not work with multi-measure sensors on the GLX, when a probe corresponding to a prior measurement in the measurement list has been disconnected.

Affected Products:

PS-2002 Xplorer GLX
PS-2170 PASPORT Chemistry Sensor

Problem/Symptom:
Calibration does not work with multi-measure sensors on the GLX, when a probe corresponding to a prior measurement in the measurement list has been disconnected.

PASCO Solution:

This is the result of a bug that exists in GLX firmware version 1.42 and all prior versions. Any calibrateable multi-measure sensor is affected by this firmware issue.

As a workaround, connect all probes that correspond to prior measurements in the measurement list of the multimeasure sensor.