November, 2001, Biology Experiment:

Transpiration in a Plant Leaf

- Equipment
- Equipment Set-up
- Software Set-up
- Data Collection Procedure
- Data Analysis
- Data Table
- Conclusions and Extensions




Purpose:

In this experiment, students study the movement of water within a plant, measuring a change in pressure as the plant transpires.

Equipment:

Back to top

Equipment Setup:

  1. Put the barb end of a quick-release connector into one end of the plastic tubing. Use a drop of glycerin for lubrication if needed.



  2. Cut the stem of the plant seedling 2-3 cm above the soil. Immediately submerge the cut end into water, and leaving it submerged, shave the freshly cut end to a 45-degree angle.

  3. Fill the tubing with water so that there is a 2-3 cm air gap at one end: raise the end with the quick-release connector and allow excess water to spill out the other end.

  4. Placing your thumb over the free end of the tubing, put the tubing under water and insert the cut plant stem into the tubing. Avoid creating any air bubbles in the tubing; pull the tubing away from the stem if air bubbles form.

  5. Spread the petroleum jelly around the end of the tube to create an airtight seal between the top edge of the plastic tubing and the plant stem.

  6. Secure the plant in an upright position with the clamp to the lab stand.

  7. Mount the Pressure Sensor to the lab stand’s support rod using the other clamp. The pressure port should be 5-7 cm above the cut end of the plant stem to prevent water from entering the Pressure Sensor.


  8. Align the quick-release coupling on the end of the plastic tubing with the pressure port of the Pressure Sensor. Push the coupling onto the port, and then turn the coupling clockwise until it clicks (about 1/8 of a turn). Make sure no water enters the sensor.

Back to top

Software Setup:

  1. PASPORT users: Connect the pressure sensor / plant assembly to the USB link or Xplorer; choose "Launch DataStudio". A graph display (Pressure vs. Time) will appear.

  2. Double-click the Digits display icon, or drag it to the Relative Pressure data channel, to open a Digits display of Pressure (kPa).

    Users of ScienceWorkshop 500 or 750 interface click here.

Back to top

Data Collection Procedure:

  1. Click the Start button ( ) to begin collecting data.

  2. Collect data for approximately 10 minutes, then click the Stop ( ) button.
    Note: If the pressure does not change or starts to increase, it is likely that your system has developed a leak. Try reseating the plant in the tube and applying more petroleum jelly around the end of the tube to create a seal.

  3. Set up the electric fan at least one meter away from the plant. Turn the fan on to a low setting so that it blows a light breeze over the plant.

  4. Record another data run, again for approximately 10 minutes.

Back to top

Data Analysis:

  1. Scale the axes to fit the data using the Scale to Fit button ( ) in the Graph toolbar.

  2. Use the Smart Tool ( ) to find the pressure (P1) at 20 seconds (t1) and the pressure (P2) at 475 seconds (t2) for Run #1 without the fan, and record the values in the data table below.

  3. Repeat the above process for Run #2 with the fan.

  4. Calculate the change in time and change in pressure for each data run, and record the values in the data table.

  5. Divide P by t to calculate the transpiration rate for each run.

Data Table:

Transpiration
Run
P1
P2
P (kPa)
t1
t2
t (s)
Rate of Transpiration
(kPa/min)
No Fan
 
 
 
 
 
 
 
With Fan
 
 
 
 
 
 
 

Conclusions and Extensions:

  1. What was the rate of pressure change in the tubing? How does the change in pressure correspond to an increase or decrease in water loss through the seedling’s stomates? Explain.

  2. Did the fan affect the rate of pressure change? Explain how the fan affects transpiration. What natural phenomenon does the fan mimic?

  3. Describe some adaptations that plants have to enable them to minimize water loss from their leaves.